Comparison of stereotactic plans for brain tumors with two different multileaf collimating systems

Livia Marrazzo 1,, Margherita Zani 2, Stefania Pallotta 1,2, Daniela Greto 3, Silvia Scoccianti 3, Cinzia Talamonti 1,2, Giampaolo Biti 2,3, Marta Bucciolini 1,2
PMCID: PMC5711251  PMID: 24423831

Abstract

Linac‐based stereotactic radiosurgery (SRS) has been widely used for treating small intracranial lesions. This technique allows conforming the dose distribution to the planning target volume (PTV), providing a steep dose gradient with the surrounding normal tissues. This is realized through dedicated collimation systems. The present study aims to compare SRS plans with two collimating systems: the beam modulator (BM) of the Elekta Synergy linac and the DirexGroup micromultileaf collimator (μMLC). Seventeen patients (25 PTVs) were planned both with BM and μMLC (mounted on an Elekta Precise linac) using the Odyssey (PerMedics) treatment planning system (TPS). Plans were compared in terms of dose‐volume histograms (DVH), minimum dose to the PTV, conformity index (CI), and homogeneity index (HI), as defined by the TPS, and doses to relevant organs at risk (OAR). The mean difference between the μMLC and the BM plans in minimum PTV dose was 5.7%±4.2% in favor of the μMLC plans. No statistically significant difference was found between the distributions of the CI values for the two planning modalities (p=0.54), while the difference between the distributions of the HI values was statistically significant (p=0.018). For both BM and μMLC plans, no differences were observed in CI and HI, depending on lesion size and shape. The PTV homogeneity achieved by BM plans was 15.1%±6.8% compared to 10.4%±6.6% with μMLC. Higher maximum and mean doses to OAR were observed in the BM plans; however, for both plans, dose constraints were respected. The comparison between the two collimating systems showed no substantial differences in terms of PTV coverage or OAR sparing. The improvements obtained by using μMLC are relatively small, and both systems turned out to be adequate for SRS treatments.

PACS numbers: 87.53.Ly, 87.55.dk, 87.56.nk

Keywords: central nervous system tumors, stereotactic radiosurgery, multileaf collimators, conformity index, homogeneity index

I. INTRODUCTION

Linac‐based stereotactic radiosurgery (SRS) is extensively used for treating small intracranial lesions. One of the main advantages exploited by this technique is the ability to conform the dose distribution to the planning target volume (PTV), providing a steep dose gradient between the PTV and the surrounding normal tissues. This is realized through the use of dedicated collimation systems with small size leaves. Commercially available micromultileaf collimators (μMLC) with narrow leaf widths are generally “add‐on” devices for use on nondedicated linac units.

Linacs with thin MLC leaf widths (4 to 5 mm at isocenter) are now routinely available. The possibility to achieve clinically acceptable treatment plans for SRS without the need of any additional devices, which require extra commissioning and an increase in treatment setup time, should be investigated.

Since 2001, at the Radiotherapy Unit of Florence University Hospital, the DirexGroup (Wiesbaden, Germany) μMLC, mounted on an Elekta (Stockholm, Sweden) Precise linac, has been used for SRS plans. About 200 patients have been treated for intracranial lesions using this device. Main advantages of the system are the small leaf size and the bidirectional leaf setting. On the other hand, the collimator needs to be manually fixed to the linac gantry prior to each usage, and it reduces the clearance around the patient, limiting the possible beam entrances.

In 2008 an Elekta Synergy linac with beam modulator (BM) was installed for treating small lesions and for SRS. This collimator does not need any manual fixation and optimizes the clearance around the patient.

It was shown by Bortfeld et al. (1) that for a 6 MV photon beam the optimal leaf width according to basic physics is in the range of 1.5‐2 mm. Therefore an evaluation of the potential of the two MLCs is meaningful in order to assess their performances for SRS plans.

The present study aims to compare SRS plans obtained with the two available collimating systems in order to estimate their advantages and limitations. The comparison presented in this work is purely a computer‐based planning study and it does not aim to evaluate the dose distributions actually delivered by the two systems.

II. MATERIALS AND METHODS

A. Patients

The latest seventeen patients (for a total of 25 PTVs) treated with SRS for brain tumors at the Radiotherapy Department of the Florence University were planned both with the BM of the Elekta Synergy linac and the μMLC mounted on the Elekta Precise linac, using the treatment planning system (TPS) Odyssey (PerMedics Inc., San Bernardino, CA, version 4.4 and 4.6), a system capable of managing a collimator with bidirectional leaf banks.

Odyssey TPS uses for the three‐dimensional dose distribution computation the method of lateral scatter convolution. Dose is computed at each point within a 3D fan grid containing the volume of tissue irradiated by the beam. The 3D fan grid is constructed from a stack of divergent 2D grid planes perpendicular to the beam direction. The number of stack planes is set to 32, 64, or 128, according to whether the speed category is fast, medium, or slow, respectively. Both electron and photon beams are supported by the system. (2)

For the purposes of this work, a calculation grid with a voxel size of 0.65 mm was adopted. No inverse optimization was used for plans calculation.

In Table 1, the clinical sample characteristics are summarized (number of lesions per patient, lesion's shape, and anatomical location). For each patient the radiation oncologist identifies the gross tumor volume (GTV) on the CT. Clinical target volume (CTV) is obtained by a 1 mm isotropic expansion of GTV; PTV is obtained by a 1 mm isotropic expansion of CTV

The median and mean PTV volumes were 5.1 cm3 and 6.2 cm3 (range 0.62‐20.7 cm3), respectively. These volumes correspond to a median and mean diameter (obtained by approximating the volume to a sphere) of 2.1 cm and 2.2 cm (range 1.1‐3.4 cm), respectively.

Doses ranged from 10 Gy to 24 Gy and were delivered in a single fraction. The median number of beam entrances per treatment was 11 (range 8‐18).

All patients were immobilized for CT scanning and treatment using the HeadFIX frame (Elekta, Stockholm, Sweden) with an individualized vacuum cushion and a patient‐specific dental mold. This assures accuracy in patient repositioning of about 1 mm, as reported in the literature. 3 , 4

Table 1.

Main features of the lesions of the selected sample

Patient Lesion Anatomical Location Shape Distance from the Fixation Point (dental mold) (cm)
1 1 Right parietal lobe Spherical 17
2 1 Cerebellar vermis Ovoidal 12
3 1 Right parietal lobe Spherical 17
4 1 Left thalamus Spherical 14
5 1 Left temporal lobe Spherical 18
2 Right occipital lobe Spherical 16
3 Right parietal lobe Spherical 14
4 Left parietal lobe Spherical 18
6 1 Right parietal lobe Spherical 14
7 1 Left cerebellar lobe Spherical 14
8 1 Left frontal lobe Spherical 12
2 Right parietal lobe Spherical 17
9 1 Right temporal lobe Irregular shape 11
10 1 Left parietal occipital lobe Spherical 16
11 1 Right parietal temporal lobe Ovoidal 14
12 1 Left temporal lobe Ovoidal 12
13 1 Right occipital lobe Ovoidal 18
2 Left frontal lobe Spherical 15
14 1 Left occipital lobe Irregular shape 19
15 1 Left occipital lobe Spherical 15
2 Right parietal lobe Bilobed 17
16 1 Right temporal lobe Ovoidal 12
2 Left frontal lobe Spherical 11
17 1 Right frontal lobe Spherical 10
2 Left cerebellar lobe Spherical 12

Patients were imaged on a Brilliance Big Bore CT (Philips Healthcare, Andover, MA), with the following image acquisition parameters: 512×512 slice resolution (0.68 mm pixel size), slice thickness 2 mm.

Patient positioning before treatment was verified with a couple of portal images (antero‐posterior and latero‐lateral).

B. Collimating systems characteristics

The μMLC consists of two levels of thin leaves (48 pairs), able to move along two orthogonal directions. The exclusive feature of bidirectional leaf setting allows one to shape irregular small fields with great accuracy. The 14 inner pairs of each level are made of thinner leaves. They form a fine resolution field of about 50mm×45mm at the isocenter. The maximum square field that can be defined at isocenter is 97mm×108mm. The μMLC is 540 mm in outer diameter, 135 mm in height, and 31 kg in weight, and must be manually mounted on the linac prior to each usage. (5)

Elekta Synergy BM consists of 40 pairs of individually controlled leaves with no backup jaws. The small size treatment head maximizes clearance around the patient and contributes to the wide variety of possible treatment approaches, including noncoplanar treatments. (6) In Table 2, the two MLC system characteristics are compared.

Table 2.

Main features of the two MLC systems

μMLC
BM
Leaf settings Two orthogonal leaf banks One leaf bank
Leaves number 48 pairs 40 pairs
Physical leaf width (mm) 2.1 (14 inner pairs), 3.6 (10 outer pairs) 1.7
Isocenter leaf width (mm) 3.2 (14 inner pairs), 5.5 (10 outer pairs)a 4
3.6 (14 inner pairs), 6.2 (10 outer pairs)b
Max. field size (mm2)
97×108
210×160
Leaf thickness (mm) 37.5 75
Max. leakage between leaves (%) 5 2
Transmission (%) 0.4 1
Distance from the source (cm)
65
a
39
58
b
Linac head clearance (cm) 31.7 45
a

a For the bank further away from the radiation source.

b

b For the bank closer to the radiation source.

C. Plan comparison

The same beam entrances were used for both plans. We tried to maintain as much as possible the parameters of the plans in order to better point out the different behaviors of the two collimating systems. For each beam entrance, the collimator rotation ensuring the better conformation was chosen. Margins to the target were the same for both plans, since they are fixed by our internal protocol in SRS. The plans obtained with the two collimating systems were normalized to the same dose at the isocenter. This is one of the possible approaches, sometimes proposed in literature. (7) Plans were then compared by visual inspection of isodose lines, by dose‐volume histogram (DVH) analysis, and by using the PTV coverage, the homogeneity index (HI), and the conformity index (CI) as defined by the TPS Odyssey. Unlike the Radiation Therapy Oncology Group (RTOG) definitions, (8) HI was defined by the TPS as the ratio of the maximum PTV dose (Dmax) to the minimum PTV dose (Dmin):

HI=DmaxDmin (1)

CI was defined as the ratio of the volume contained within the smallest isodose cloud that covers the target (VSI) to the PTV volume (TV). Once the minimum PTV dose matches the prescription dose, these definitions coincide with the RTOG ones.

CI=VSITV (2)

The PTV coverage was determined as the ratio of Dmin to the prescribed dose (PD).

PTV coverage=DminPD (3)

PTV coverage homogeneity was evaluated by calculating the relative difference:

PTV coverage homogeneity=(DmaxDmin)PD (4)

It should be noted that this parameter actually represents an estimate of the inhomogeneity of target coverage, since the higher the value of PTV coverage homogeneity, the less homogeneous the coverage.

The minimum doses have been obtained as the lowest doses to the PTV. We decided to not consider an eventual single voxel, at the PTV edge, outside the highest isodose curve encompassing the PTV. These protruding PTV voxels would affect the minimum PTV dose with modest clinical significance. Maximum doses are assessed as the ones received by the 1% of PTV volume.

Analysis was also conducted splitting PTVs into two classes according to size (“small volumes” were considered the ones 6cm3; “big volumes” the ones >6cm3, with 6 cm3 being the average PTV size in our sample).

Organs at risk (OAR) were compared in terms of maximum and mean dose.

III. RESULTS

A. Analysis on PTVs

When looking at the PTV coverage, plans obtained with μMLC achieved better results. The minimum PTV dose was 91.3%±6.5% (±1SD) (range 80.1%‐98.6%) and 95.2%±6.6% (±1SD) (range 78.9%‐99.8%) for the BM and μMLC plans, respectively. The average difference between the μMLC and the BM plans in minimum PTV dose was 5.7%±4.2% in favor of the μMLC plans.

Concerning the CI, the average value was 2.1±0.5 (±1SD) (range 1.5–3.3) and 2.0±0.4 (±1SD) (range 1.5–3.0) for the BM and μMLC plans, respectively. No statistically significant difference was found between the distributions of the CI values for the two planning modalities (p=0.54). In Fig. 1, the values obtained with BM plans are plotted versus the values obtained with μMLC. The linear fit to the experimental data is quite near to the quadrant bisector, indicating only a slight difference.

Concerning the HI, the average value was 1.2±0.1 (±1SD) (range 1.1‐1.4) and 1.1±0.1 (±1SD) (range 1.0‐1.4) for the BM and μMLC plans, respectively. The difference between the distributions of the HI values for the two planning modalities was statistically significant (p=0.018). In Fig. 2, the values obtained with BM plans are plotted versus the values obtained with μMLC. In this case, the linear fit to the experimental data is quite far from the quadrant bisector, in favor of the μMLC plans.

PTV coverage homogeneity was 15.1%±6.8% for BM compared to 10.4%±6.6% of μMLC, showing a more homogeneous dose distribution for μMLC plans. Results are summarized in Table 3.

For both BM and μMLC plans, no differences were observed in CI and HI, depending on lesion size (data not shown). In the same way, no differences were observed depending on lesion shape (spherical versus differently shaped PTVs).

Figure 1.

Figure 1

Conformity index of the BM plans versus the μMLC plans. The dashed line is the quadrant bisector.

Figure 2.

Figure 2

Homogeneity index of the BM plans versus the μMLC plans. The dashed line is the quadrant bisector.

Table 3.

PTVs analysis results

μMLC
BM
PTV coverage (%) (±1SD)
95.2±6.6
91.3±6.5
CI (±1SD)
2.0±0.4
2.1±0.5
HI (±1SD)
1.1±0.1
1.2±0.1
PTV coverage homogeneity (%) (±1SD)
10.4±6.6
15.1±6.8

B. organs at risk

In Figs. 3 and 4, maximum and mean doses to main OAR are reported for the two plan modalities. OAR doses were always slightly higher with BM plans compared to μMLC plans. However, for both plans, all the OAR doses were below their respective maximum clinical doses.

Figure 3.

Figure 3

Maximum doses for selected OAR for BM and μMLC plans.

Figure 4.

Figure 4

Mean doses for selected OAR for BM and μMLC plans.

IV. DISCUSSION

This paper deals with the comparison between SRS plans obtained with two different collimating systems and may be of interest to those acquiring dedicated MLC equipment for SRS or owning a small‐width MLC linac.

Several papers in the literature approached the issue of comparing different collimating systems for SRS. It has been found that a linac‐based SRS with multiple static beams allows better normal tissue sparing for irregularly shaped targets than multiple arcing beams using circular collimators. 9 , 10 , 11 Lead alloy blocks provide an accurate conformation to the target, but are unpractical and need to be customized. The introduction of the MLC has simplified the radiotherapy process, although the target conformity can be reduced by the leaves size.

Adams et al. (12) compared SRS treatment plans using a 10 mm leaf width MLC and blocks and found that, for a similar PTV coverage, a significantly higher volume of normal tissue was irradiated to >50% and >80% dose using the MLC.

Kubo et al. (13) undertook a study to analyze the impact of collimator leaf width on SRS using 1.7 or 3.0 mm leaf μMLCs. They found, by analyzing CI, isodose distributions, and dose‐volume histograms, that the use of both μMLCs allows one to meet the RTOG guidelines for stereotactic radiosurgery.

Monk et al. (14) dosimetrically compared the BrainLAB m3 μMLC (minimum leaf width of 3 mm) (BrainLAB, Feldkirchen, Germany) with the Varian 2100EX 5 mm leaf MLC (Varian Medical Systems, Palo Alto, CA) for SRS treatment of intracranial lesions. They found an improved CI for the 3 mm leaf width (m3) (1.5±0.2 vs. 1.6±0.2). Concerning the surrounding tissue sparing, when expressed as the maximum dose received by the adjacent critical normal tissue, there was no statistically significant difference. When expressed as the mean difference in the volume of critical structure encompassed by the 50% and 70% isodose levels, the values were 5.7% and 4.9% favoring the m3. The authors conclude that the clinical importance of these small differences is difficult to assess and individual centers may question their choice of equipment when outfitting a SRS service.

Similar conclusions were found by Tanyi et al. (15) who analyzed the impact of two MLC systems (2.5 mm vs. 5 mm leaves) for linear accelerator‐based intracranial SRS. They demonstrated on 68 lesions a small dosimetric advantage of the 2.5 mm leaf width MLC system in terms of dose conformation median reduction, of normal tissue exposure, and steepness of the dose falloff. The authors pointed out that these improved dosimetric results may not translate into clinical benefits.

In the paper by Dhabaan et al., (16) the performance of a high‐definition multileaf collimator of 2.5 mm leaf width (MLC2.5) was compared to standard 5 mm leaf width MLC (MLC5) for the treatment of intracranial lesions using dynamic conformal arcs technique. The MLC2.5 shows a dosimetric advantage over the MLC5, both in treatment conformity and normal tissue sparing. The advantages are more evident when target shape complexity increases.

Nill et al. (17) proposed a planning study to analyze the impact of different leaf widths on the achievable dose distributions with intensity‐modulated radiation therapy. They found the MLC with the smallest leaf width always yields the best PTV coverage. Reducing the leaf width from 4 to 2.75 mm results in a slight enhancement of the PTV coverage but in no significant improvement for most OARs. The disadvantage of the reduction of the leaf width is the increasing number of segments due to the more complex fluence patterns and, therefore, increased delivery time.

In a recent study on a new add‐on device (2.5 mm leaves), Godwin and colleagues (18) investigated a range of mechanical and dosimetric characteristics which included inter‐and intraleaf leakage, light/radiation field congruence, leaf position reproducibility, radiation penumbra, total scatter factors, and mechanical rotational stability.

Asnaashari et al. (19) compared the dosimetric parameters of two multileaf collimator systems, namely the BM and Radionics micro‐MLC (Radionics Inc., Burlington, MA), using measurements and Monte Carlo simulations.

In our analysis, planning outcome is evaluated in terms of PTV coverage and dose to the OAR. Concerning PTV coverage, when looking at the CI, no significant differences were observed between the two collimating systems. μMLC plans presented some advantages in terms of HI and minimum PTV dose. However, differences are relatively small and the planning outcomes with respect to PTV between the two systems are comparable. There is no obvious clinical advantage in using the μMLC compared to the BM collimator. No dependence on tumor size and shape was observed. It must be taken into account that our sample is mainly constituted of regularly shaped targets.

When looking at the OAR, slight higher maximum and mean doses are observed in the BM plans compared to the μMLC plans. This can be partly related to the higher transmission under the BM leaves due to the absence of backup jaws. Again the clinical significance of these data is debatable, since for both plans maximum OAR doses were acceptable and below the clinical threshold limit.

The measured 80%‐20% penumbra (P2080) for the two MLCs is summed in Table 4. Evaluation of the beam penumbra demonstrated a larger P2080 by up to about 1.5 mm for BM compared with the μMLC. These differences are mainly related to the different features of the two collimators leaf tip ends and their distances to the source, and they will probably have less impact on a three‐dimensional dose distribution for a multiple beam arrangement when the contribution from all beams is considered. Nonetheless, looking at the measured profiles for a 4×4cm2 field (Fig. 5(a)) and to those extracted by a 2D dose map for both plans (Fig. 5(b)), a difference can be observed, being the μMLC profile is flatter and characterized by a smaller penumbra, compared to BM.

Table 4.

P2080 for the two MLCs

Field Size (cm) Depth (cm) P2080 (mm)
μMLC
2.4 1.5 2.8
8 1.5 3.2
2.4 10 3.3
8 10 4.6
BM 2.4 1.5 3.9
8 1.5 4.4
2.4 10 4.1
8 10 6.3

Figure 5.

Figure 5

Comparison of measured profiles for (a) a μMLC and a BM 6 MV, 4×4cm2 field. Comparison of profiles extracted from a TPS map for (b) a μMLC plan and a BM plan. The horizontal line marks the half maximum dose.

V. CONCLUSIONS

In a group of 25 lesions treated with SRS, the comparison between the two collimating systems μMLC and BM showed no substantial differences in terms of PTV coverage or OAR sparing. In particular, no significant differences were observed between the two collimating systems in terms of CI, while μMLC plans presented some advantages in terms of HI and minimum PTV dose, even though both BM and μMLC plans provided acceptable HI.

Higher maximum and mean doses to OAR were observed in the BM plans compared to the μMLC plans but, for both plans, maximum OAR doses were acceptable and below the recommended constraints. The targets we examined were mainly spherically shaped, so that these findings cannot be generalized to irregular lesions. In these conditions, when viewed quantitatively, the improvements obtained by using μMLC are relatively small and both systems turned out to be adequate for SRS treatments.

ACKNOWLEDGMENTS

Authors wish to acknowledge Dr. Nicolò Bechini for the help in data collection.

REFERENCES

  • 1. Bortfeld T, Oelfke U, Nill S. What is the optimum leaf width of a multileaf collimator. Med Phys. 2000;27(11):2494–501. [DOI] [PubMed] [Google Scholar]
  • 2. PerMedics Inc. Odyssey User Manual. Document no. P3000520. San Bernardino, CA: PerMedics Inc.; 1999. [Google Scholar]
  • 3. Sweeney R, Bale R, Vogele M, et al. Repositioning accuracy: comparison of a non‐invasive head holder with thermoplastic mask for fractionated radiation therapy and a case report. Int J Radiation Oncology Biol Phys. 1998;41(2):475–83. [DOI] [PubMed] [Google Scholar]
  • 4. Olch AJ and Lavey RS. Reproducibility and treatment planning advantages of a carbon fiber relocatable head fixation system. Radiother Oncol. 2002;65(3):165–68. [DOI] [PubMed] [Google Scholar]
  • 5. Bucciolini M, Russo S, Banci Buonamici F, Pini S, Silli P, Dosimetric characterization of a bi‐directional micromultileaf collimator for stereotactic applications. Med Phys. 2002;29(7):1456–63. [DOI] [PubMed] [Google Scholar]
  • 6. Patel I, Glendinning AG, Kirby MC. Dosimetric characteristics of the Elekta Beam Modulator. Phys Med Biol. 2005;50(23):5479–92. [DOI] [PubMed] [Google Scholar]
  • 7. Khoo VS, Oldham M, Adams EJ, Bedford JL, Webb S, Brada M. Comparison of intensity‐modulated tomotherapy with stereotactically guided conformal radiotherapy for brain tumors. Int J Radiat Oncol Biol Phys. 1999;45(2):415–25. [DOI] [PubMed] [Google Scholar]
  • 8. Shaw E, Scott C, Souhami L, et al. Single dose radiosurgical treatment of recurrent previously irradiated primary brain tumors and brain metastases: final report of RTOG protocol 90‐05. Int J Radiat Oncol Biol Phys. 2000;47(2):291–98. [DOI] [PubMed] [Google Scholar]
  • 9. Hamilton RJ, Kuchnir FT, Sweeney P, et al. Comparison of static conformal field with multiple noncoplanar arc techniques for stereotactic radiosurgery or stereotactic radiotherapy. Int J Radiat Oncol Biol Phys. 1995;33(5):1221–28. [DOI] [PubMed] [Google Scholar]
  • 10. Laing RW, Bentley RE, Nahumb AE, Warrington AP, Brada M. Stereotactic radiotherapy of irregular targets: a comparison between static conformal beams and non‐coplanar arcs. Radiother Oncol. 1993;28 (3):241–46. [DOI] [PubMed] [Google Scholar]
  • 11. Kubo HD, Pappas CTE, Wilder RB. A comparison of arc‐based and static mini‐multileaf collimator‐based radiosurgery treatment plans. Radiother Oncol. 1997;45(1):89–93. [DOI] [PubMed] [Google Scholar]
  • 12. Adams EJ, Cosgrove VP, Shepherd SF, et al. Comparison of a multi‐leaf collimator with conformal blocks for the delivery of stereotactically guided conformal radiotherapy. Radiother Oncol. 1999;51(3):205–09. [DOI] [PubMed] [Google Scholar]
  • 13. Kubo HD, Wilder RB, Pappas CTE. Impact of collimator leaf width on stereotactic radiosurgery and 3D conformal radiotherapy treatment plans. Int J Radiat Oncol Biol Phys. 1999;44(4):937–45. [DOI] [PubMed] [Google Scholar]
  • 14. Monk JE, Perks JR, Doughty D, Plowman PN. Comparison of a micro‐multileaf collimator with a 5‐mm‐leaf‐width collimator for intracranial stereotactic radiotherapy. Int J Radiation Oncology Biol Phys. 2003;57(5):1443–49. [DOI] [PubMed] [Google Scholar]
  • 15. Tanyi JA, Kato CM, Chen Y, Chen Z, Fuss M. Impact of the high‐definition multileaf collimator on linear accelerator‐based intracranial stereotactic radiosurgery. Br J Radiol. 2011;84(1003):629–38. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16. Dhabaan A, Elder E, Schreibmann E, et al. Dosimetric performance of the new high‐definition multileaf collimator for intracranial stereotactic radiosurgery. J Appl Clin Med Phys. 2010;11(3):197–211. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17. Nill S, Tuecking T, Muenter MW, Oelfke U. Intensity modulated radiation therapy with multileaf collimators of different leaf widths: a comparison of achievable dose distributions. Radiother Oncol. 2005;75(1):106–11. [DOI] [PubMed] [Google Scholar]
  • 18. Godwin GA, Simpson JB, Mugabe KV. Characterization of a dynamic multi‐leaf collimator for stereotactic radiotherapy applications. Phys Med Biol. 2012;57(14):4643–54. [DOI] [PubMed] [Google Scholar]
  • 19. Asnaashari K, Chow JC, Heydarian M. Dosimetric comparison between two MLC systems commonly used for stereotactic radiosurgery and radiotherapy: a Monte Carlo and experimental study. Phys Med. 2013;19(4):350–56. [DOI] [PubMed] [Google Scholar]

Articles from Journal of Applied Clinical Medical Physics are provided here courtesy of Wiley

RESOURCES