TABLE III.

Inter–observer reproducibility

Left Ventricle Right Ventricle

Global Bland–Altman
Bias ± Limits
CoV Bland–Altman
Bias ± Limits
CoV

Circum. Strain (%) 0.5 ± 2.2 3.6 −0.5 ± 2.4 3.1
Long. Strain (%) 0.8 ± 2.7 4.8 0.7 ± 2.1 3.3
Radial Strain (%) 1.3 ± 8.4 7.1
Torsion (°) 0.1 ± 0.8 2.5 −0.4 ± 3.0 14.8
1 Dyssynchrony (%) 0.3 ± 1.6 5.6 −0.1 ± 2.2 9.8

Segmental

Circum. Strain (%) 0.3 ± 3.6 5.7 −1.1 ± 6.2 9.8
Long. Strain (%) 0.6 ± 4.6 7.7 0.3 ± 6.9 10.3
Radial Strain (%) −1.0 ± 17.6 12.0
Torsion (°) 0.1 ± 4.4 16.8 −0.5 ± 8.9 44.4
1,2 Delay Times (%) 0.2 ± 4.0 0.2 ± 4.0
1

Expressed as a percent of the cardiac cycle

2

Negative is early contraction; Positive is late contraction