Published in final edited form as: IEEE Trans Med Imaging. 2016 Dec 29;36(5):1076–1085. doi: 10.1109/TMI.2016.2646321

TABLE IV.

Inter–test reproducibility

Left Ventricle Right Ventricle

Global Bland–Altman
Bias ± Limits
CoV Bland–Altman
Bias ± Limits
CoV

Circum. Strain (%) 0.7 ± 1.8 3.8 0.4 ± 1.5 2.8
Long. Strain (%) 0.0 ± l.3 2.4 −0.4 ± 3.1 4.8
Radial Strain (%) 0.6 ± 11.5 11.6
Torsion (°) 0.2 ± 1.2 5.6 0.6 ± 2.0 8.4
1 Dyssynchrony (%) 0.4 ± 3.6 11.8 0.3 ± 2.2 10.8

Segmental

Circum. Strain (%) 0.5 ± 4.0 6.6 0.8 ± 4.6 7.1
Long. Strain (%) −0.1 ± 4.2 7.6 −0.3 ± 6.3 9.2
Radial Strain (%) 1.1 ± 21.3 18.0
Torsion (°) 0.2 ± 4.7 19.6 0.6 ± 8.2 28.0
1,2 Delay Times (%) −0.8 ± 5.9 0.4 ± 5.6
1

Expressed as a percent of the cardiac cycle

2

Negative is early contraction; Positive is late contraction