Table 3. QCA analysis.

Genoss DES™ (n=38) Promus Element™ (n=39) p
RVD (mm)
Pre-procedure 3.24±0.41 3.12±0.43 0.199
Post-procedure 3.30±0.30 3.33±0.30 0.652
9-month follow-up 3.17±0.35 3.16±0.39 0.932
Minimal lumen diameter (mm)
In-stent
Pre-procedure NA NA NA
Post-procedure 2.95±0.35 3.03±0.30 0.269
9-month follow-up 2.84±0.42 2.87±0.56 0.743
In-segment
Pre-procedure 0.83±0.29 0.88±0.45 0.572
Post-procedure 2.90±0.27 3.00±0.36 0.159
9-month follow-up 2.79±0.40 2.85±0.56 0.615
DS %
In-stent
Pre-procedure NA NA NA
Post-procedure 10.6±3.48 9.16±4.85 0.140
9-month follow-up 14.18±9.31 13.73±14.30 0.871
In-segment
Pre-procedure 74.41±8.62 72.65±12.14 0.465
Post-procedure 12.10±3.93 10.12±4.94 0.056
9-month follow-up 15.46±9.39 14.54±14.08 0.738
Acute gain (mm)
In-stent 1.79±0.40 1.82±0.36 0.742
In-segment 1.74±0.39 1.78±0.38 0.579
Late loss (mm)
In-stent 0.11±0.25 0.16±0.43 0.567
In-segment 0.11±0.26 0.15±0.43 0.558
Lesion length (mm) 23.8±8.1 22.8±4.9 0.531
Restenosis 1 (2.6) 2 (5.1) 1.000

Data are shown as mean±standard deviation or number (%).

DES = drug-eluting stent; DS = diameter stenosis; NA = not available; QCA = quantitative coronary angiography; RVD = reference vessel diameter.