Skip to main content
. 2013 Jan 7;14(1):147–160. doi: 10.1120/jacmp.v14i1.4032

Figure 5.

Figure 5

The capability ratio (Cp) and acceptability ratio (Cpk) of first run and longest run for output constancy check measured by central ionization chamber of RBA‐3 device with different time to calculate the control limits for (a) 6 MV photon beams and (b) 12 MeV electron beams.