Table 1.

The number of first run points before out‐of‐control limits and number of longest run points on the X‐charts for 6 and 10 MV photon beams, and 6, 9, 12, 16, and 20 MeV electron beams. Each month is equal to 4 data points. N is the number of data points used to calculate the limits.

N 6X '09 10X '09 6E '09 9E '09 12E '09 16E '09 20E '09
(months)
1st
Long
1st
Long
1st
Long
1st
Long
1st
Long
1st
Long
1st
Long
1 5 5 5 7 5 5 8 11 13 29 6 12 5 11
2 12 15 12 15 14 15 14 15 13 29 14 15 14 15
3 12 15 14 15 14 15 14 15 13 29 14 15 14 15
4 30 30 30 30 30 30 14 15 13 29 14 15 14 29
5 30 30 30 30 14 15 12 15 12 29 14 15 14 15