Skip to main content
. 2013 Jan 7;14(1):258–272. doi: 10.1120/jacmp.v14i1.4037

Table 1.

Values and uncertainty estimates (1 σ) for Method 1, irradiation in air. Dwell times were calculated to deliver 1 Gy. The measured dose was 1.022 Gy (i.e., 2.2% higher than predicted). The uncertainty in d is estimated to be 0.5 mm, resulting in a 2% uncertainty in the calculated absorbed dose. The total uncertainty is the sum in quadrature of all the components and does not include the uncertainty in the TLD reading (0.7%).

Quantity Value Units Uncertainty (%)
Calculation of
Dw,inair
from AKR
AKR (at 1 m)
8.85×106
Gys1
1
Distance d in air 50 mm 2
Correction for catheter attenuation, Attcath 1.0017 0.2
Correction for TLD holder attenuation, AttHold 1.013 2
Ratio of mass energy absorption coefficients, (μ¯trρ)aw 1.11 3
Dwell time for 1 Gy = 258.2 s 4%
Attcath×AttHold/[AKR×(1000/d)2×(μ¯σρ)aw]
Measurement with ionisation chamber
Mu
1.81×108
C < 0.1
kTPkelec
1.006 0.1
NK,ARP
4.89×107
GyC1
1
Ratio of mass energy absorption coefficients, (μ¯trρ)aw 1.11 3
kn
1.035 0.4
Distance d in air 50 mm 2
Dw,inair=Mu×kTPkelec×NK,ARP×(μ¯trρ)aw×kn
1.022 Gy 4%