Table 3.

Values and uncertainty estimates (1 σ) for Method 2, irradiation in the Nucletron jig in water. Dwell times were calculated to deliver 1 Gy. The measured dose was 1.011 Gy (i.e., 1.1% higher than predicted). The uncertainty in d is estimated to be 0.5 mm, resulting in a 2% uncertainty in the calculated absorbed dose.

Quantity Value Units Uncertainty in
Dw
(%)
Calculation of
Dw
from AKR
AKR (at 1 m)
8.85×106
Gys1
1
Distance d in water 50 mm 2
Dose at 50 mm per unit air kerma at 1 m, R 446 2
Dwell time for 1 Gy=1/(R×AKR) 253.3 s 3%
Measurement with ionisation chamber
Mu
1.81×108
C < 0.1
kTPkelec
1.006 0.1
ND calculated by TRS‐277 from Nk,Ir192
4.74×107
GyC1
1
Average stopping power ratio, water to air, s 1.137 0.3
pWall
0.99 0.5
pcel
0.98 0.5
pn(=kn)
1.035 1
Distance d in water 50 mm 2
Dw=Mu×kTPkelec×ND×s×pnwall×pncel×pn
1.011 Gy 3%