. 2013 Jan 7;14(1):273–288. doi: 10.1120/jacmp.v14i1.4077

Table 3.

Energy match analysis between TrueBeam, Clinac, and Trilogy linear accelerators.

Energy (MV) Data Field Size
(cm×cm)
TrueBeam Clinac Trilogy
4×4
1.37 1.38 1.41
dmax (cm)
10×10
1.41 1.41 1.44
40×40
1.32 1.20 1.29
4×4
61.4 61.4 61.7
6 PDD10 (%)
10×10
66.1 66.2 66.1
40×40
71.8 71.3 71.5
4×4
0.625 0.627 0.626
TPR20/10
10×10
0.666 0.667 0.668
40×40
0.745 0.741 0.746
4×4
2.83 3.00 2.80
dmax (cm)
10×10
2.78 2.60 2.59
40×40
2.02 2.15 2.08
4×4
74.9 75.1 74.8
15 PDD10 (%)
10×10
76.7 76.6 76.5
40×40
76.9 77.3 77.5
4×4
0.731 0.731 0.733
TPR20/10
10×10
0.763 0.762 0.763
40×40
0.810 0.807 0.811