Skip to main content
. 2013 May 6;14(2):172–185. doi: 10.1120/jacmp.v14i2.4136

Table 4.

Mean ± 1 SD statistics for beam delivery and verification using binned dose rate and continuously variable dose rate (CVDR).

Agility a MLCi
Binned Dose Rate CVDR Binned Dose Rate
Delivery time (s)
171.4±10.4
132.6±7.4
281.8±100.0
Verification gamma (3%/3 mm)
95.0±2.9
95.2±2.8
96.2±1.3
a

Items in bold are statistically significant at the p < 0.05 level, compared against the MLCi.