Water mean CT value and standard deviation versus dose and SAFIRE strength (Fig. 1) |
120 kV, 300/200/100/50 mAs, 1.0 sec rotation time, collimation, 2 mm slice thickness, pitch 1.0 |
FBP:B31s SAFIRE: I31s S1, S2, S3, S4, S5 |
30 cm diameter water‐filled phantom |
Water standard deviation versus different reconstruction kernels (Table 2) |
120 kV, 200 mAs, 1.0 sec rotation time, collimation, 2 mm slice thickness, pitch 1.0 |
FBP:B31s, B36s, B40s, B70s SAFIRE: I31s, I36s, I40s, I70h S3 |
30 cm diameter water‐filled phantom |
Noise power spectrum NPS (Fig. 2) |
120 kV, 200 mAs, 1.0 sec rotation time, collimation, 2 mm slice thickness, pitch 1.0 |
FBP: B40s SAFIRE: I40s S1, S2, S3, S4, S5 |
30 cm diameter water‐filled phantom |
CT number linearity and CT number accuracy (Table 3) |
120 kV, 200 mAs, 1.0 sec rotation time, collimation, 2 mm slice thickness, pitch 1.0 |
FBP:B31s SAFIRE: I31s S1, S3, S5 |
Catphan 600, sensitometry module |
Low contrast resolution (Fig. 3, Table 4) |
120 kV, 300/200/100 mAs, 1.0 sec rotation time, collimation, 2 mm slice thickness, pitch 1.0 |
FBP : B31s SAFIRE: I31s S1, S2, S3 |
Catphan 600, low‐contrast resolution module |
Transverse spatial resolution and modulation transfer function MTF (Fig. 4) |
120 kV, 200/50 mAs, 1.0 sec rotation time, collimation, 2 mm slice thickness, pitch 1.0 |
FBP: B70s SAFIRE: I70h S1, S3, S5 |
Catphan 600, high‐resolution module |
Coronal spatial resolution (Fig. 5) |
120 kV, 500 mAs, 1.0 sec rotation time, collimation, pitch 1.0 |
FBP: B46f SAFIRE: I46f S5 MPR: 1 mm recon thickness, 0.1 mm image increment |
3D spatial resolution phantom |