Skip to main content
. 2013 Jul 8;14(4):206–215. doi: 10.1120/jacmp.v14i4.4313

Table 4.

Reproducibility results for distance‐to‐agreement measurements in three irradiations of the phantom using a linac‐based delivery system

Inferior Edge Superior Edge
Film Plane DTA (mm) Std. Dev. (mm) DTA (mm) Std. Dev. (mm)
Sagittal 1.2 1.1 1.3 1.0
Coronal 1.2 1.1 1.2 1.0