TABLE V.

Uncertainty assessment for the experimental dosimetry. (N: normal distribution)

Contributions Distr. Std. Unc.
Rats 900MHz Mice 1900MHz
E/H field measurement N 12.8% 11.3%
Field control N 4.7% 4.7%
Chamber temperature fluctuations N 9.6% 9.6%
Temperature probe accuracy N 1.2% 1.2%
Combined standard uncertainty k=1 16.7% 15.6%
Expanded uncertainty k=2 33.5% 31.2%