Skip to main content
. 2013 Sep 6;14(5):52–67. doi: 10.1120/jacmp.v14i5.4253

Table 4.

Dose increase factor (DIF) assigned to neutron contamination in a 15 MV photon beam produced by four prostheses: Co‐Cr‐Mo, Stainless steel, Ti‐alloy, and Ti in a water phantom

Dose Increase Factor (DIF) ± Uncertainty
Depth in Phantom (cm) Co‐Cr‐Mo Stainless Steel Ti‐alloy Ti
1
1.05±0.11
1.02±0.10
1.04±0.10
1.06±0.10
2
1.04±0.10
1.00±0.09
1.01±0.09
1.04±0.10
3
1.03±0.10
0.97±0.10
0.96±0.09
1.01±0.10
4
0.99±0.11
0.92±0.10
0.90±0.09
0.96±0.10
Water 5
1.05±0.12
0.93±0.11
0.84±0.10
0.98±0.11
6
1.71±0.19
1.37±0.17
0.99±0.12
1.09±0.14
7
5.85±0.61
4.58±0.55
1.90±0.23
2.17±0.27
8
36.85±4.12
26.27±3.16
8.05±0.96
9.09±1.08
9
184.55±21.80
137.33±18.75
40.66±4.93
43.17±5.20
11
2197.63±298.29
1593.77±228.52
468.48±63.86
524.81±71.4
Prosthesis 12
1229.37±171.61
925.15±132.63
277.89±38.96
311.25±43.58
13
1729.75±247.97
1347.23±105.44
422.54±40.84
471.48±67.80
15
138.49±10.99
110.58±16.92
36.86±3.06
40.77±3.37
16
31.05±4.79
23.49±3.62
8.49±1.36
9.49±1.51
17
8.84±1.39
6.68±1.20
2.25±0.38
2.63±0.44
18
2.86±0.53
2.17±0.42
0.93±0.19
1.15±0.23
Water 19
0.94±0.19
0.88±0.17
0.58±0.13
0.79±0.18
20
0.39±0.08
0.71±0.16
0.60±0.12
0.49±0.11
21
0.40±0.10
0.71±0.18
0.46±0.12
0.49±0.13
22
0.83±0.26
0.45±0.13
0.49±0.15
0.46±0.15