Skip to main content
. 2013 Sep 6;14(5):52–67. doi: 10.1120/jacmp.v14i5.4253

Table 5.

Dose increase factor (DIF) assigned to total dose (neutron dose+photon dose) in a 15 MV photon beam produced by four prostheses: Co‐Cr‐Mo, Stainless steel, Ti‐alloy, and Ti in a water phantom

Dose Increase Factor (DIF) ± Uncertainty
Depth in Phantom (cm) Co‐Cr‐Mo Stainless Steel Ti‐alloy Ti
1
1.00±0.08
1.00±0.08
1.00±0.08
1.00±0.08
2
0.99±0.08
1.00±0.08
0.99±0.08
1.01±0.08
3
1.01±0.10
0.99±0.09
1.00±0.09
0.99±0.09
4
1.00±0.02
1.00±0.02
0.99±0.02
1.00±0.02
Water 5
1.00±0.02
1.00±0.02
1.00±0.02
1.00±0.02
6
1.00±0.02
1.00±0.02
1.00±0.02
1.00±0.02
7
1.00±0.02
1.00±0.02
1.00±0.02
1.00±0.02
8
1.00±0.02
1.00±0.02
1.00±0.02
1.00±0.02
9
1.00±0.02
1.00±0.02
1.01±0.02
1.00±0.02
11
0.83±0.02
0.86±0.02
0.96±0.02
0.85±0.02
Prosthesis 12
0.68±0.01
0.75±0.02
0.88±0.02
0.79±0.02
13
0.54±0.01
0.63±0.01
0.81±0.02
0.70±0.02
15
0.44±0.01
0.55±0.01
0.69±0.02
0.68±0.02
16
0.43±0.01
0.53±0.01
0.69±0.02
0.67±0.02
17
0.43±0.01
0.52±0.01
0.68±0.02
0.67±0.02
18
0.43±0.01
0.52±0.01
0.68±0.02
0.67±0.02
Water 19
0.43±0.01
0.52±0.01
0.68±0.02
0.66±0.02
20
0.43±0.01
0.52±0.01
0.68±0.02
0.66±0.02
21
0.43±0.01
0.52±0.01
0.67±0.02
0.66±0.02
22
0.43±0.01
0.52±0.01
0.67±0.02
0.66±0.02