Table 3.

Application of failure mode and effects analysis for the preplanning imaging and volume determination stage in tomotherapy. Failure modes having an assigned RPN>80 are reported

Subprocess No Potential Failure Mode Potential Causes of Failure Potential Effects of Failure O S D RPN
VIII Positioning/alignment of the patient on the CT couch on the basis of the laser 1 Wrong definition of the isocenter Inadequate CT laser alignment Systematic shift of the patient position 4 8 3 96
X Reference markers positioning to indicate previous treatment fields 2 Missing marker positions Lack of attention or incomplete compilation of the RT record Previous treatment not taken into consideration/suboptimal planning 4 6 3 96
XXIII Registration of possible other images from different techniques and check of consistency 3 Wrong registration Consistency not verified Wrong PTV and OAR definition 3 7 5 105
XXIV Definition of the GTVs and/or CTVs contour on the basis of the anatomical and/or functional information (BTV) 4 Wrong CTV definition Lack of attention/inadequate skill Wrong dose distribution 3 7 5 105
XXV Definition of ITV in the case of 4D CT acquisition 5 Incorrect ITV construction Lack of attention/inadequate skill Wrong PTV 3 6 5 90
XXVI Automatic contour of the OAR 6 Missing OAR definition Lack of attention/inadequate skill Unintended normal tissue irradiation 3 8 5 120
XXVII Manual contour of the OAR 7 Missing OAR definition Lack of attention/inadequate skill Unintended normal tissue irradiation 3 8 5 120
XXVIII Defnition of the planning structure at risk through automatic expansion of OAR 8 Wrong expansion Lack of attention/inadequate skill Unintended normal tissue irradiation 3 5 6 96
XXIX Automatic expansion with definition of the CTVs and PTVs margins 9 Wrong expansion Lack of attention/inadequate skill Unintended normal tissue irradiation 3 5 6 96
XXX Definition and contouring of overlapping regions 10 Wrong/missed definition Lack of attention/inadequate skill or not detailed information on previous treatment Unintended normal tissue irradiation or wrong dose distribution 4 7 5 140