Skip to main content
. 2013 Sep 6;14(5):13–24. doi: 10.1120/jacmp.v14i5.3828

Table 3.

Translational and rotational setup errors result in a decrease in the pass rate of γ analysis with various criteria on the average and standard deviation

ArcCHECK
Delta4
Setup Error
3%/3mm
3%/2mm
2%/2mm
3%/3mm
3%/2mm
2%/2mm
Translation
1 mm
0.7±1.0
1.4±1.5
2.0±2.0
0.6±0.6
1.6±1.5
2.1±2.2
2 mm
2.8±1.7
5.0±2.9
6.7±3.5
2.3±1.9
5.4±3.3
7.3±4.2
3 mm
6.9±3.3
10.0±4.9
13.2±5.5
6.1±3.8
11.1±5.5
14.6±6.7
Rotation
1.8±1.9
3.7±2.9
4.9±3.7
0.5±0.8
1.5±1.5
2.5±2.6
8.6±4.7
12.1±5.5
14.5±6.6
2.4±1.7
5.0±2.9
7.0±3.7