Skip to main content
. 2013 Nov 8;14(6):5–17. doi: 10.1120/jacmp.v14i6.4358

Table 2.

Backscatter factors (BSF) measured locally compared to IPEM code of practice data.( 2 , 16 )

Tube Potential (kV) Applicator Size (cm) Measured BSF IPEM BSF Difference (%)
60 3.0 1.132 1.114 +1.6
10.0 1.185 1.179 +0.5
120 3.0 1.143 1.164 −1.8
10.0 1.309 1.335 −1.9
160 10.0 1.369 1.350 +1.4
220 10.0 1.296 1.292 +0.3