Estimate of BSF measurement uncertainties using EBT2 film

Component Uncertainty (%)
Output stability 0.5
Film positioning 1.0
Film nonuniformity 1.5
OD variation over pixels 1.5
Scanner variation 1.5
Film calibration 1.0
Water equivalence of phantom 1.0
Total 3.2