Table 4.

Summary of reported BSF measurements using novel detectors

Study Machine Beam Quality (HVL) Detector Reference Data Maximum Difference to Reference Data Measurement Uncertainty
Butson and Cheung(3)
Gulmay D3300 50–150 kV (1.4mmAl0.6mmCu) GAFCHROMIC EBT IPEM 1.8% Not given
Kim et al.(4)
Pantak DXT300 50,100,280 kV (1.5mmAl3.2mmCu) GAFCHROMIC EBT AAPM 2% 2.5%
Smith et al.(5)
Pantak DXT300 50,100,280 kV (1.5mmAl3.2mmCu) GAFCHROMIC EBT2 AAPM 3.5% 4.0%
Nelson and Hill(17)
Pantak Therapax (DXT) 300 75–300 kV (2.2mmAl3.2mmCu) LiF TLD AAPM & IPEM
~6%
1.1%4.7%
Mart et al.(6)
Gulmay D3100 20, 100 kV (0.08, 4.0 mm Al) GAFCHROMIC EBT2 AAPM a IPEM 11%, 7%
2%7%
OSLD AAPM a IPEM 14%
1%10%
This study (2013) Gulmay D3225 60–220 kV (1.3mmAl2.2mmCu) GAFCHROMIC EBT2 IPEM 1.9% 3.2%
a

Originally, data for 20 kV were compared to a very old publication (BJR supplement 10, 1961) which has since been updated three times and currently (BJR supplement 25, 1996) agrees with AAPM code of practice data within 1%. Therefore current AAPM code of practice values are used for this comparison.