. 2012 Jan 5;13(1):148–163. doi: 10.1120/jacmp.v13i1.3585

Table 1.

Accuracy of the energy determination by the step‐wedge procedure. Changes in beam energy were invoked by changing the Injector I. The ratio D20/D10 is measured using ionization chambers in solid water and using the signals registered by the MVCT detector for the different step‐wedge levels.

Injector I (Vdc) Solid Water D20/D10 Step Wedge Difference (%)
3.10 0.535 0.536 0.2
3.96 0.521 0.520
0.2
4.07 0.519 0.519 0.0
4.26 0.516 0.514
0.4
5.00 0.502 0.499
0.6