. 2012 Jan 5;13(1):44–57. doi: 10.1120/jacmp.v13i1.3347

Figure 6.

Figure 6

The effect of shifting the isocenter of either the RPO field (a) or the LAO field (b); shifts are perpendicular to the match plane. The dose difference is measured using a Farmer chamber (accurate to within ± 0.5% and positioned to within ±0.2mm) for the OMF technique only.

The discrepancies between calculation and measurement are attributed to partial volume effects in the ion chamber. A reduction in sensitivity to target motion in the direction of the shift is realized when the OOMF or SJOMF techniques are employed.