Skip to main content
. 2012 Mar 8;13(2):55–72. doi: 10.1120/jacmp.v13i2.3725

Table 4.

Binned DRs of the linac tested. Nominal DRs are suggested for entry into the TPS, and are used in the text and figures.

Binned DR Nominal DR (MU/min) 6 MV Actual DR (MU/min) 10 MV Actual DR (MU/min)
Max 600 435 475
Max/2 300 216 236
Max/4 150 108 118
Max/8 75 53 59
Max/16 38 27 30
Max/32 19 12 14