Skip to main content
. 2012 Mar 8;13(2):55–72. doi: 10.1120/jacmp.v13i2.3725

Table 6.

Maximum values of DR, MLC speed, backup jaw speed, and GS determined using internal consistency tests described in text.

DR Accuracy a MLC
(Δx/Δt)maxcm/sec
Jaw
(Δx/Δt)maxcm/sec
Gantry
(Δθ/Δt)max/sec
Nominal 2.0 1.5 6.0
6 MV 0.99 1.99 1.49 6.00
10 MV 0.99 1.94 1.46 5.90
a

Displayed (MU/min)/Measured (MU/min); error of 0.5sec overlaps with 1.00