. 2012 Jul 5;13(4):141–151. doi: 10.1120/jacmp.v13i4.3796

Figure 9.

Figure 9

The results from the deformable registration tests. The upper panels show an example head and neck patient CBCT to FBCT deformable image registration with our extended FOV CBCT images: (a) the images before applying the deformable registration, with the target (CBCT) image, the source (FBCT) image, and a color overlay (yellow indicates good agreement, green and red indicate negative and positive disagreement) of the two images; (b) the deformed target, the source, and the overlay images, after applying the deformable registration. The lower panels show the registration between a CBCT and its artificially deformed counterpart, (c) before the registration and (d) after the registration. The CBCT image was used as the target, and the artificially deformed image with a Gaussian was used as the source.