Skip to main content
. 2012 May 10;13(3):124–151. doi: 10.1120/jacmp.v13i3.3729

Figure 10.

Figure 10

Localization accuracy of ETX and CBCT: (a) hidden target test error distribution on AP and LAT portal images (units in mm), (b) hidden target test error distribution of ETX with systematic offsets removed (μ±σ=0.0±0.3mm for L/R, 0.0±0.1mm for P/A, and 0.1±0.2mm for S/I), and (c) angular discrepancy between 3D/2D ETX and 3D/3D CBCT image registration algorithms (left) and robotic couch error estimated from 8 embedded BBs on 9 pairs of pre/ post CBCT images (right).