Table 2.
The results of the flattened and FFF beams, where diff (%) = (FFF‐Flat)/Flat.
|
|
|||||||
|---|---|---|---|---|---|---|---|
| Farmer Chamber | PinPoint Chamber | Plane‐parallel Chamber | |||||
| SAD (cm) | 100 | 150 | 100 | 150 | 100 | 150 | |
| 6X Flat | 1.004 | 1.000 | 1.000 | 1.000 | 1.018 | 1.015 | |
| 6X FFF | 1.008 | 1.003 | 1.006 | 1.000 | 1.014 | 1.012 | |
| Diff (%) | 0.4% | 0.3% | 0.6% | 0 | ‐0.4% | ‐0.3% | |
| 10X Flat | 1.012 | 1.000 | 1.005 | 1.000 | 1.011 | 1.008 | |
| 10X FFF | 1.015 | 1.008 | 1.005 | 1.000 | 1.019 | 1.014 | |
| Diff (%) | 0.3% | 0.8% | 0 | 0 | 0.8% | 0.6% | |