Skip to main content
. 2012 Sep 6;13(5):140–154. doi: 10.1120/jacmp.v13i5.3837

Table 2.

Ion chamber measures to validate independent dose calculation technique.

Eclipse Calculated Dose / Pinnacle Calculated Dose (Avg ± STD) Measured / Pinnacle Calculated Dose (Avg ± STD) MapCHECK Gamma Passing Rate (Avg ± STD) %
Treatment Site IC1 IC2 IC1 IC2
Head‐and‐ Neck
0.972±0.005
0.980±0.009
1.022±0.008
1.013±0.010
96.3±1.62
Breast
0.990±0.004
0.996±0.01
1.010±0.006
1.009±0.020
94.8±1.13
Prostate
0.995±0.01
1.01±0.004
0.994±0.009
0.997±0.012
98.2±0.58
Rectum
0.975±0.008
1.005±0.01
1.017±0.016
1.017±0.025
94.9±1.68