. 2012 Sep 6;13(5):108–116. doi: 10.1120/jacmp.v13i5.3382

Table 6.

Gamma analysis and point dose results of IMRT and VMAT plans of AAPM TG 119 test cases. Gamma criteria of 3% dose difference (DD) and 3 mm distance to agreement (DTA).

VMAT IMRT
Point Dose (Gy) Gamma Point Dose (Gy) Gamma <1(%)
Test Case Measured Planned ratio <1(%) Measured Planned ratio <1(%)
Prostate 1.989 2.066 0.98 96.79 2.0 2.0072 0.996 97.67
Head & Neck 1.915 1.943 1.014 98.15 1.960 1.984 0.988 95.92
C‐shaped 2.263 2.292 0.987 96.86 2.301 2.325 0.987 96.11
Multi Target 2.301 2.280 1.009 98.99 2.245 2.288 0.981 97.92