. 2012 Sep 6;13(5):117–125. doi: 10.1120/jacmp.v13i5.3766

Table 1.

Mean deviation of MVision positioning accuracy at six known positions for the different CBCT imaging MUs, reconstruction slice thicknesses, and planning CT slice thicknesses for the phantom study.

Deviation (mm)
CBCT Imaging MUs a CBCT Slice Thickness b (mm) Planning CT Slice Thickness c (mm) SI LR AP
5 1 1
0.6±0.5
0.4±0.3
0.5±0.0
3
1.0±0.0
0.8±0.5
0.4±0.5
8 1 1
0.6±0.5
0.5±0.4
0.5±0.0
3
1.3±0.5
1.0±0.7
0.3±0.5
5 3 1
0.5±0.4
0.5±0.4
0.5±0.0
3
1.3±1.0
0.6±0.5
0.4±0.5
8 3 1
0.5±0.6
0.5±0.4
0.5±0.0
3
1.0±0.8
0.5±0.4
0.3±0.5
a

Compared with CBCT imaging MUs: FSI=0.0, p=1.0; FLR=0.037, p=0.849; FAP=0.261, p=0.614.

b

Compared with CBCT slice thickness: FSI=0.090, p=0.767; FLR=0.926, p=0.346; FAP=0.0, p=1.0

c

Compared with planning CT slice thickness: FSI=7.254, p=0.013; FLR=0.037, p=0.849; FAP=2.348, p=0.139