Skip to main content
. 2012 Sep 6;13(5):198–214. doi: 10.1120/jacmp.v13i5.3856

Table 3.

Differences in correction factors between central and off‐axis detectors for various gantry angles. Difference (%)=((offaxisdetector'sCF(θ)centraldetector'sCF(θ))/centraldetector'sCF(θ))×100.

Photon Energy 6 MV 10 MV
Gantry angle, θ (degrees) Average Range Average Range
0°–70° 0.4
1.0 to 2.2
0.6
1.0 to 2.4
80°–110° 1.1
6.1 to 7.0
1
3.7 to 6.0
120°–180° 0.1
2.1 to 2.0
0.6
1.3 to 1.9