Table 4.

Comparison in average passing rates (%) between iPlan MC‐calculated and MatriXX dose distributions without and with “central correction” or “entire correction” at θ=0°180° for 10×10cm2, 15×15cm2, and 20×20cm2 fields with (a) 6 MV and (b) 10 MV photons. Gamma evaluation was performed with 2% dose difference and 2 mm distance‐to‐agreement (DTA) criteria.

(a).
Angular Correction
Field Size (cm 2 ) No Correction Central Correction Entire Correction
10×10
64.7 91.4 93.1
15×15
63.4 92.0 94.4
20×20
58.8 90.1 94.3
(b).
Angular Correction
Field Size (cm 2 ) No Correction Central Correction Entire Correction
10×10
64.3 91.7 93.0
15×15
60.3 93.3 94.6
20×20
56.7 93.8 95.6