Table 5.

Comparison in passing rates (%) between iPlan MC‐calculated and MatriXX dose distributions without and with “central correction” or “entire correction” for various simple plans in a 10×10cm2 field of 6 MV photons. Gamma evaluation was performed with 2% dose difference and 2 mm DTA, and 3% dose difference and 3 mm DTA.

Angular Correction
No Correction Central Correction Entire Correction
Plan No. 2%/2 mm 3%/3 mm 2%/2 mm 3%/3 mm 2%/2 mm 3%/3 mm
1 57.4 71.8 90.1 99.5 90.7 99.7
2 90.0 99.3 81.3 97.8 90.7 99.6
3 88.4 98.8 92.1 99.3 94.5 99.3
4 68.2 93.8 95.5 99.7 95.6 99.7
5 80.4 96.9 95.2 99.2 95.2 99.2
6 78.9 96.6 96.2 99.9 96.2 99.9