Skip to main content
. 2012 Sep 6;13(5):198–214. doi: 10.1120/jacmp.v13i5.3856

Table 6.

Comparison in passing rates (%) between TPS‐calculated and MatriXX dose distributions without and with “central correction” or “entire correction” for IMRT (iPlan MC) and VMAT (Eclipse AAA) plans. Gamma evaluation was performed with 2% dose difference and 2 mm DTA, and 3% dose difference and 3 mm DTA.

Angular Correction
No Correction Central Correction Entire Correction
Plan No. 2%/2 mm 3%/3 mm 2%/2 mm 3%/3 mm 2%/2 mm 3%/3 mm
1 (VMAT) 90.7 97.7 91.3 98.3 91.5 98.6
2 (VMAT) 92.9 99.1 93.4 99.1 93.9 99.1
3 (IMRT) 87.9 97.9 94.8 99.8 94.9 99.8
4 (IMRT) 65.8 87.2 87.0 99.2 87.1 99.2
5 (IMRT) 60.4 77.5 86.9 94.6 87.7 95.4
6 (IMRT) 82.8 99.2 95.1 99.6 95.6 99.6