Skip to main content
. 2012 Nov 8;13(6):303–317. doi: 10.1120/jacmp.v13i6.3992

Table 3.

Percentage of pixels passing the 5%/3 mm criteria in the moving target and cord cylinder for various motion patterns and dose calculation methods.

Motion Pattern Film Orientation Percentages of Pixels Passing 5%/3 mm
Target
M_1
Coronal
94±8
Transverse
89±12
Mean±SD
92±5
M_2
Coronal
91±8
Transverse
92±10
Mean±SD
91±4
M_3
Coronal
95±7
Transverse
100±0
Mean±SD
98±4
M_4
Coronal
94±7
Transverse
99±1
Mean±SD
96±3
M_5
Coronal
99±1
Transverse
100±0
Mean±SD
99±0
M_6
Coronal
96±6
Transverse
100±0
Mean±SD
97±3
M_7
Coronal
86±6
Transverse
97±5
Mean±SD
91±1
Overall
Mean±SD
95±7
Off‐target OAR
M_5
Coronal
99±2
M_6
Coronal
88±8
M_7
Coronal
85±15
Overall
Mean±SD
91±10