Skip to main content
. 2012 Nov 8;13(6):34–45. doi: 10.1120/jacmp.v13i6.3760

Table 4.

Comparison of total MUs, PTV coverage, and plan QA for prostate, hypoprostate, and hypolung SmartArc plan generated using binned and continuous dose rate mode (control point spacing=6deg, no leaf speed constraint). Interp. Plan refers to QA plans interpolated to 2 degree control point spacing.

ArcCheck
Treatment Site Dose Rate Maximum Delivery Time (s) Total MU % PTV coverage Ori. Plan Interp. Plan
Prostate Binned 90 453 93.3 89.3 91.5
Continuous 90 447.5 92.6 89.9 92.6
Hypoprostate Binned 90 822.3 92.3 88.5 92.3
Continuous 90 855.2 92.2 88.8 92.9
Hypolung Binned 300 2808.5 95.1 92.4 94.3
Continuous 300 2858 95.1 92.5 94.6