Skip to main content
. 2012 Nov 8;13(6):112–125. doi: 10.1120/jacmp.v13i6.4007

Table 2.

Values of the conformity index (CI) and the uniformity (UPTV) determined for each case in the database when planned using the standard method, iterative reoptimization, and using noncoplanar beams. CI is unitless while UPTV is in units of Gy. Note that lower values of both CI and UPTV are considered superior.

Patient CI
UPTV
Standard Method Iterative Reoptimization Noncoplanar Beams Standard Method Iterative Reoptimization Noncoplanar Beams
1 1.59 1.19 1.41 25.04 15.39 28.95
2 1.39 1.25 1.30 13.94 13.28 16.03
3 1.12 1.24 1.08 24.53 25.7 14.58
4 1.31 1.13 1.53 17.33 15.81 17.44
5 1.38 1.36 1.34 19.85 13.08 21.69
6 1.40 1.23 1.24 18.94 16.14 17.69
7 1.16 1.52 1.24 21.13 27.8 23.07
8 1.21 1.35 1.17 10.9 10.63 15.72
9 1.44 1.34 1.45 14.94 10.18 15.34
10 1.61 1.22 1.55 20.24 17.31 19.89
11 1.63 1.31 1.58 29.24 14.73 28.2
12 2.24 1.81 2.47 28.33 20.12 27.23
13 1.26 1.14 1.21 15.04 12.96 15.55
14 1.37 1.13 1.47 19.24 10.37 21.45
15 1.48 1.47 1.19 13.89 14.73 14.38