. 2010 Oct 27;12(1):218–233. doi: 10.1120/jacmp.v12i1.3279

Table 5.

Dosimetric and geometrical parameters used for Diso reconstruction.

Beam Gantry Angle
ρ(g/cm3)
ρiso(g/cm3)
z (cm)
ziso(cm)
w (cm)
wiso(cm)
L (cm)
Diso,TPS(cGy)
Diso(cGy)
1 0.926 0.961 23.4 16.6 21.7 16.0 9.5 128
126.0±3.0
2 130° 0.921 1.050 23.4 10.5 21.6 11.0 9.5 82
82.8±1.5
3 200° 0.947 0.785 25.2 7.2 23.9 5.7 8.8 66
64.1±1.1

Notes: ρ and ρiso are the physical density along beam axis across the patient and up to isocenter, respectively; z and ziso are the geometrical patient thickness and the isocenter depth along beam axis; w and wiso are the radiological paths along the beam axis; L is the equivalent square field size, Diso,TPS is the calculated dose by TPS at isocenter and Diso is the mean measured dose.