Figure 3.

Dose profiles in the x‐axis at depths of 5 (a), 10 (b) and 15 cm (c) in the solid water phantom as shown in Fig. 1. Doses in the profiles are calculated by the AAA, CCC and MC simulations using the 6 MV photon beams () at gantry angle 0°. All profiles are normalized to the prescription points in Fig. 1.