. 2010 Nov 3;12(1):108–121. doi: 10.1120/jacmp.v12i1.3424

Table I.

Percentage differences of the mean dose and SDs calculated for depth ranges of 5–20 (gantry angles equal to 0° and 5°) or 5–15 cm (gantry angle equal to 45°) using the 6 and 15 MV photon beams with the calculated configuration as in Fig. 1. A negative value in the table means an underestimation of dose by the AAA and CCC compared to the MC. The field sizes are 4×4 (6 MV), 7×7 (15 MV), 10×10 and 20×20cm2 and gantry angles are 0°, 5° and 45°.

4×4 or 7×7cm2
10×10cm2
20×20cm2
Energy 45° 45° 45°
AAA% CCC% AAA% CCC% AAA% CCC% AAA% CCC% AAA% CCC% AAA% CCC% AAA% CCC% AAA% CCC% AAA% CCC%
6 MV
10.5±1.3
3.4±0.9
15.1±3.6
3.7±1.5
6.4±0.6
21.3±0.5
7.6±2.6
2.1±1.3
3.2±1.2
1.8±3.8
9.0±4.5
18.3±7.0
16.3±2.1
6.7±2.1
3.7±2.7
3.2±1.6
17.6±2.3
12.5±2.4
15 MV
7.8±1.6
4.6±1.7
12.0±3.5
7.6±2.0
4.0±0.7
9.1±0.9
5.5±1.2
1.7±1.4
0.4±2.2
0.6±0.4
13.5±2.8
19.6±2.1
18.0±1.3
8.3±1.8
4.0±1.7
3.9±1.7
22.6±3.0
21.4±3.9