Table 1.

Radial dose function for point source approximation for an active source without the applicator and a single active pellet in positions 1(gp1) to 6(gp6) inside the applicator.

r (cm)
gP
pellet source without the applicator
gp1
gp2
gp3
gp4
gp5
gp6
0.5 1.007 1.012 1.013 1.013 1.012 1.011 1.012
1 1 1 1 1 1 1 1
1.5 0.994 0.996 0.993 0.995 0.995 0.995 0.995
2 0.998 0.997 0.992 0.989 0.989 0.992 0.990
2.5 0.991 0.984 0.983 0.983 0.986 0.984
3 0.985 0.984 0.974 0.979 0.979 0.978 0.981
3.5 0.979 0.977 0.968 0.980 0.979 0.972 0.979
4 0.968 0.971 0.958 0.971 0.970 0.964 0.971
5 0.956 0.960 0.953 0.948 0.949 0.953 0.948
6 0.943 0.943 0.939 0.943 0.939 0.939 0.938
7 0.927 0.926 0.923 0.924 0.919 0.922 0.917
8 0.911 0.907 0.894 0.909 0.905 0.899 0.905
9 0.893 0.886 0.886 0.873 0.876 0.883 0.878
10 0.870 0.865 0.858 0.842 0.850 0.860 0.851