Skip to main content
. 2011 Apr 14;12(3):142–160. doi: 10.1120/jacmp.v12i3.3475

Table 2.

Average radiation penumbra (80%–20%) width of 1×20cm2 field measured at different off‐axis distances for 6 MV X‐rays at 1.5 cm and 10 cm depth.

Average Beam Penumbra (SD) in cm at Off‐axis Distances
Depth of Measurement 0 cm 4 cm 8 cm 12 cm
1.5 cm 0.17 (0.01) 0.16 (0.01) 0.16 (0.01) 0.16 (0.01)
10 cm 0.37 (0.03) 0.39 (0.04) 0.38 (0.03) 0.38 (0.03)