Table 6.

Quantification of the effect of TPS angle discretization (2°, 4° and 6° angular CP spacing) when estimating continuous, full rotation simple 2 cm‐wide arc. All values are in passing rates (%, calculated vs. calculated) with global normalization dose equal to the maximum dose in the corresponding subvolume and a lower threshold dose of 10% of the normalization dose.

100% Norm
6020
4020
ROI Analyzed Dose (Gy)
γ(3/3)1
γ(2/2)1
γ(1/1)1
γ(3/3)1
γ(2/2)1
γ(1/1)1
1 cm Diameter Cylinder 4.375 100 100 94.4 100 100 100.0
3 cm Diameter Cylinder 4.375 100 99.9 84.6 100 100 96.7
5 cm Diameter Cylinder 4.375 100 100 94.0 100 100 98.9
10 cm Diameter Cylinder 4.375 100 100 98.4 100 100 99.7
ArcCHECK Detector Surface 0.430 56.5 51.3 42.3 70.2 58.5 43.5