Table 1.

The differences between expected absolute doses (predicted by the TPS) and absolute doses measured in our ion chamber phantom (second and third columns) for each patient's treatment plan. The shapes of the dose distributions were verified based on the number of diodes satisfying MapCHECK's gamma criterion (fourth column). Finally, we compare the MapCHECK dose on the central axis to its predicted value (fifth and sixth columns). Summary statistics of these measures are given in the bottom rows.

Absolute Dose Mapcheck
Patient Expected Measured Passing Measured Expected
1 228.9 224.2 100 78.3 75.85
2 235.8 235.8 100 92.33 97.45
3 206.2 205.4 97.9 121.81 121.62
4 391.5 380.5 98.5 124.45 124.75
5 232.6 238 96.9 158.78 161.95
6 254.6 250.4 99.5 119.87 122.52
7 241.1 242.3 95.2 92.86 76.1
8 200.7 195.4 96 82.86 75.42
9 214 217.9 97 82.98 81.27
10 215.5 209.3 96.3 48.1 44.96
11 160.4 163.1 95.8 92.7 93.73
12 203.2 204.3 97.6 108.21 108.43
13 283.6 282.3 99.7 284.35 297.97
14 281.8 279.4 95.5 80.26 78.99
15 236.2 227.1 97.2 127.1 113.93
16 359.6 354.7 96.3 45.01 32.85
17 445.9 442.4 98.5 75.86 73.21
MEAN 258.3294 256.0294 97.52353 106.8135 104.7647
S.D 72.62554 70.79058 1.562626 52.48433 57.05988