Figure 3

Distribution of measured gap errors for six patients treated with supine technique. Measurements done after planned couch shifts with BB placement described in Section II: a) gap 1: mean error=1.0mm(σ=1.3mm); b) gap 2: mean error=1.6mm(σ=2.1mm).