k0 and ks factors determined for 15 beams (supplied by three linacs) of the 38 wedged beams examined in this work.

Clinac 2100 Varian Elekta Precise Siemens Oncor
Nominal Mega Voltage (MV) 6 10 15 6 10 15 6 10 15
Wedge Angle (°) 15 60 15 60 15 60 60 60 60 15 60 15 60 15 60
k0
1.63 3.13 1.42 2.67 1.33 2.53 3.01 2.89 2.90 1.77 3.44 1.85 3.25 1.43 2.45
ks106(cCU/a.u)