. 2011 Nov 15;12(4):15–23. doi: 10.1120/jacmp.v12i4.3554

Table 1.

The direct posterior (gantry angle 180°), maximum, and average beam attenuations (in percentage) measured with eight different couch inserts with energies of 6 and 15 MV. The average beam attenuation was calculated between gantry angles of 100°–180°. The couch inserts are in descending order with respect to attenuation with a direct posterior 6 MV beam.

6 MV 15 MV
GA 180° Max. Average (100°–180°) GA 180° Max. Average (100°–180°
BrainLAB 3.6 8.7 5.0 2.4 5.9 3.3
Qfix kVue Standard 2.1 5.2 2.9 1.4 3.3 1.9
MEDTEC 1.9 7.1 2.9 1.5 5.0 2.1
Varian Exact IGRT 1.9 4.7 2.4 1.3 3.1 1.6
DIGNITY AirPlate 1.9 3.6 2.3 1.2 2.4 1.5
Universal Sandwich Panel 1.5 7.3 2.8 1.0 5.0 2.0
Qfix kVue DoseMax 1.3 8.1 2.2 1.0 5.4 1.4
Varian Grid Insert 0.3 10.8 1.3 0.2 7.4 0.9