. 2011 Nov 15;12(4):15–23. doi: 10.1120/jacmp.v12i4.3554

Table 2.

Measured surface doses relative to the dose maximum (Dmax) (in percentage) with different couch inserts in the beam with energies of 6 and 15 MV. Measurements were performed with 10×10cm2 and 20×20cm2 open fields with a direct posterior beam (gantry angle 180°). The corrected surface dose takes into account the perturbation effect of the used plane‐parallel ionization chamber. The couch inserts are in descending order with respect to relative surface dose.

6 MV 15 MV
10×10cm2
20×20cm2
10×10cm2
20×20cm2
Measured Surface Dose 44.3 53.4 27.6 39.1
Corrected Surface Dose 35.2 44.3 21.0 32.6
Dose at Dmax 100.0 100.0 100.0 100.0
BrainLAB 98.6 99.4 84.5 89.9
Varian Exact IGRT 90.8 94.0 69.6 77.7
Universal Sandwich Panel 90.2 93.2 69.2 76.9
MEDTEC 90.1 93.4 69.1 77.3
Qfix kVue Standard 88.5 92.1 66.9 75.2
DIGNITY AirPlate 86.0 89.8 66.0 74.3
Qfix kVue DoseMax 75.1 80.9 52.8 63.0
Varian Grid Insert 61.2 69.0 41.0 51.7