Table 2.
Average transit dose rates along the x‐ and z‐axes. The average transit dose rate is also represented as a percentage of the focus dose on the day of measurement rate (2.22 Gy per minute for the 18 mm helmet). The difference between the doses measured at 0 cm along the x‐ and z‐axes results from the nonisotropic isodose distribution at isocenter; dose measured will depend on the orientation of the ionization chamber.
Average Transit Dose Rates Along the x‐axis | |||||
---|---|---|---|---|---|
Position (cm) | Measured Transit Dose Rate (cGy/min) | Percent of Focus Dose Rate (%) | |||
|
|
|
|||
|
|
|
|||
|
|
|
|||
0 |
|
|
|||
1.2 |
|
|
|||
3.2 |
|
|
|||
6.2 |
|
|
|||
Average Transit Dose Rates Along the z‐axis | |||||
Position (cm) | Measured Transit Dose Rate (cGy/min) | Percent of Focus Dose Rate (%) | |||
|
|
|
|||
|
|
|
|||
0 |
|
|
|||
1.2 |
|
|
|||
3.2 |
|
|
|||
6.2 |
|
|