. 2010 Apr 17;11(2):181–197. doi: 10.1120/jacmp.v11i2.3183

Table 1.

Nominal technical settings and measured HVLs for each imaging protocol investigated.

Scan Site XVI OBI
Head Chest Pelvis Head Chest Pelvis
kV Collimator S20 M20 M10
kV Filter a F0 F1 F1 Full bow‐tie Half bow‐tie Half bow‐tie
kVp 100 120 120 100 110 125
mA 10 40 64 20 20 80
ms/projection 10 40 40 20 20 13
# of Projections 361 643 643 360 655 655
Total mAs 36.1 1028.8 1646.1 145 262 680
Meaured HVL (mm Al) 5.9 8.9 8.9 5.4 5.7 6.4
Acquisition Angle b 350°‐190° 273°‐269° 273°‐269° 88°‐292° 88°‐92° 88°‐92°
cw cw cw cw/ccw cw/ccw cw/ccw
Acquisition Time
~70s
~120s
~120s
~30s
~60s
~60s
Axial Field of View (cm) 27 41 41 25 45 45
Long. Field of View (cm) 26 26 12.5 18 16 16
a.

On the XVI, F0 is the designation for no filter and F1 is the designation for a full bow‐tie filter.

b.

For acquisition angle, cw means clockwise and ccw means counter‐clockwise, as viewed from the patient table.